Home LiteratureArticle Details
PMID: 11300511 Published · ppublish English

Quantitation of morphine and codeine in human urine using high-filed asymmetric waveform ion mobility spectrometry (FAIMS) with mass spectrometric detection.

Journal of analytical toxicology ·Vol. 25 ·No. 2 ·2001-06-21

McCooeye M A, Ells B, Barnett D A, Purves R W, Guevremont R

Abstract

Morphine and codeine have been identified and measured in a human urine matrix using high-field asymmetric waveform ion mobility spectrometry (FAIMS) in a tandem combination with electrospray ionization (ESI) and mass spectrometric (MS) detection. The addition of helium to the nitrogen carrier gas resulted in a substantial improvement in the sensitivity of the ESI-FAIMS-MS instrument for the determination of morphine and codeine. Limits of detection in human urine were 60 ng/mL for morphine and 20 ng/mL for codeine with no clean-up, derivatization, or chromatographic separation of the sample prior to analysis.

Article Info
Journal
Journal of analytical toxicology
Abbr.
J Anal Toxicol
Published
2001-06-21
Indexed
2001-04-12
Updated
2013-11-21
Language
English
Country/Region
England
NLM ID
7705085
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]