Determination of ascorbic acid in elemental diet by high-performance liquid chromatography with electrochemical detection.

Journal of chromatography ·Vol. 606 ·No. 2 ·1992-08-14 ·Pages 277-80

Iwase H

Abstract

Determination of ascorbic acid in a multi-component elemental diet was performed by high-performance liquid chromatography with electrochemical detection. This method is suitable for the routine determination of ascorbic acid in elemental diet because it is simple, rapid, sensitive, highly selective and reproducible. The calibration graph of ascorbic acid was linear in the range 0-1.0 micrograms. The recovery of ascorbic acid was over 95% by the standard addition method. There was good agreement between the concentrations of ascorbic acid stated and found.

MeSH Terms
Ascorbic Acid/analysis Chromatography, High Pressure Liquid/methods Electrochemistry Food, Formulated/analysis
Chemicals
Ascorbic Acid
Authors & Affiliations
1 authors, click to expand affiliations / ORCID
Iwase H
Ajinomoto Co., Inc., Kawasaki, Japan.
Article Info
Journal
Journal of chromatography
Abbr.
J Chromatogr
Published
1992-08-14
Pages
277-80
Language
English
Region
Netherlands
NLM ID
0427043
Subset
IM
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]