Home LiteratureArticle Details
PMID: 14763861 Published · ppublish English Evaluation Study Journal Article Research Support, Non-U.S. Gov't Research Support, U.S. Gov't, Non-P.H.S. Validation Study

The design and fabrication of three-chamber microscale cell culture analog devices with integrated dissolved oxygen sensors.

Biotechnology progress ·Vol. 20 ·No. 1 ·2004-00-00 ·Pages 338-45

Sin A, Chin KC, Jamil MF, Kostov Y, Rao G, Shuler ML

Abstract

Whole animal testing is an essential part in evaluating the toxicological and pharmacological profiles of chemicals and pharmaceuticals, but these experiments are expensive and cumbersome. A cell culture analog (CCA) system, when used in conjunction with a physiologically based pharmacokinetic (PBPK) model, provides an in vitro supplement to animal studies and the possibility of a human surrogate for predicting human response in clinical trials. A PBPK model mathematically simulates animal metabolism by modeling the absorption, distribution, metabolism, and elimination kinetics of a chemical in interconnected tissue compartments. A CCA uses mammalian cells cultured in interconnected chambers to physically represent the corresponding PBPK. These compartments are connected by recirculating tissue culture medium that acts as a blood surrogate. The purpose of this article is to describe the design and basic operation of the microscale manifestation of such a system. Microscale CCAs offer the potential for inexpensive, relatively high throughput evaluation of chemicals while minimizing demand for reagents and cells. Using microfabrication technology, a three-chamber ("lung"-"liver"-"other") microscale cell culture analog (microCCA) device was fabricated on a 1 in. (2.54 cm) square silicon chip. With a design flow rate of 1.76 microL/min, this microCCA device achieves approximate physiological liquid-to-cell ratio and hydrodynamic shear stress while replicating the liquid residence time parameters in the PBPK model. A dissolved oxygen sensor based on collision quenching of a fluorescent ruthenium complex by oxygen molecules was integrated into the system, demonstrating the potential to integrate real-time sensors into such devices.

MeSH Terms
Animals Biomimetics/instrumentation,methods Biosensing Techniques/instrumentation,methods Cell Culture Techniques/instrumentation,methods Cell Line Equipment Design Equipment Failure Analysis Feasibility Studies Hepatocytes/physiology Humans Microfluidics/instrumentation,methods Miniaturization/methods Organ Specificity Oxygen/metabolism Oxygen Consumption/physiology Rats Respiratory Mucosa/physiology Species Specificity
Chemicals
Oxygen
Authors & Affiliations
6 authors, click to expand affiliations / ORCID
Sin Aaron
School of Chemical and Biomolecular Engineering, Cornell University, 120 Olin Hall, Ithaca, New York 14850, USA.
Chin Katherine C
Jamil Muhammad F
Kostov Yordan
Rao Govind
Shuler Michael L
Article Info
Journal
Biotechnology progress
Abbr.
Biotechnol Prog
ISSN
8756-7938
Published
2004-00-00
Pages
338-45
Language
English
Region
United States
NLM ID
8506292
Subset
IM
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]