Many microfluidic applications require the mixing of reagents, but efficient mixing in these laminar (i.e., low Reynolds number) systems is typically difficult. Instead of using complex geometries and/or relatively long channels, we demonstrate the merits of flow rate time dependency through periodic forcing. We illustrate the technique by studying mixing in a simple "T" channel intersection by means of computational fluid dynamics (CFD) as well as physically mixing two aqueous reagents. The "T" geometry selected consists of two inlet channel segments merging at 90 degrees to each other, the outlet segment being an extension of one of the inlet segments. All channel segments are 200 microm wide by 120 microm deep, a practical scale for mass-produced disposable devices. The flow rate and average velocity after the confluence of the two reagents are 48 nl s(-1) and 2 mm s(-1) respectively, which, for aqueous solutions at room temperature, corresponds to a Reynolds number of 0.3. We use a mass diffusion constant of 10(-10) m(2) s(-1), typical of many BioMEMS applications, and vary the flow rates of the reagents such that the average flow rate remains unchanged but the instantaneous flow rate is sinusoidal (with a DC bias) with respect to time. We analyze the effect of pulsing the flow rate in one inlet only as well as in the two inlets, and demonstrate that the best results occur when both inlets are pulsed out of phase. In this case, the interface is shown to stretch, retain one fold, and sweep through the confluence zone, leading to good mixing within 2 mm downstream of the confluence, i.e. about 1 s of contact. From a practical viewpoint, the case where the inlets are 180 degrees out of phase is of particular interest as the outflow is constant.

MeSH 主题词
Computer Simulation Microfluidics/instrumentation,methods Numerical Analysis, Computer-Assisted Periodicity
作者与单位
共 2 位作者,点击展开单位 / ORCID
Glasgow Ian
Department of Mechanical Engineering and New Jersey Center for Micro-Flow Control, New Jersey Institute of Technology, Newark, NJ 07102, USA.
Aubry Nadine
Article Info
Journal
Lab on a chip
Abbr.
Lab Chip
ISSN
1473-0197
Published
2003-05-00
电子出版
2003-00-30
页码
114-20
Language
English
Country/Region
England
NLM ID
101128948
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]