Pseudomonas septicemia due to deficient disinfectant mixing during reuse.

The International journal of artificial organs ·Vol. 15 ·No. 1 ·1992-01-00 ·Pages 19-24

Vanholder R, Vanhaecke E, Ringoir S

Abstract

We describe a cluster of four septicemias with pseudomonas, that occurred in a unit performing formaldehyde reuse of capillary dialyzers. Samples of blood, heparin solutions, dialysate and effluent of reused dialyzers, were evaluated bacteriologically and upon the adequacy of the reuse procedure. Pseudomonas aeruginosa, vesicularis and/or xanthomonas maltophilia were found on the blood cultures obtained during the septicemic reactions, and in the effluent of two reprocessed dialyzers not yet used (greater than 10(4) CFU/ml). These two dialyzers had also extremely low formaldehyde concentrations (0.0014 and 0.005% versus the expected 4%). Membrane and antibiogram characteristics of a Pseudomonas aeruginosa strain, recovered from the blood cultures in one patient, and of a strain found in the effluent of one of the two contaminated reprocessed dialyzers, were the same. The problem was attributed to the inadequate mixing of the disinfectant with the tap water used in the automated reprocessing device, in the absence of an alarm disclosing this failure.

MeSH Terms
Aged Bacteremia/etiology,microbiology Belgium/epidemiology Disease Outbreaks Equipment Contamination Female Formaldehyde Humans Kidneys, Artificial Male Middle Aged Pseudomonas Infections/epidemiology,etiology Renal Dialysis/instrumentation Sterilization/methods Xanthomonas/isolation & purification
Chemicals
Formaldehyde
Authors & Affiliations
3 authors, click to expand affiliations / ORCID
Vanholder R
Nephrology Department, University Hospital, Ghent, Belgium.
Vanhaecke E
Ringoir S
Article Info
Journal
The International journal of artificial organs
Abbr.
Int J Artif Organs
ISSN
0391-3988
Published
1992-01-00
Pages
19-24
Language
English
Region
United States
NLM ID
7802649
Subset
IM
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]