CFD modeling of blood flow following coil embolization of aneurysms.

Medical engineering & physics ·Vol. 26 ·No. 9 ·2004-11-00 ·页码 755-61

Byun HS, Rhee K

Abstract

In case of coil embolization of a giant or a multilobular aneurysm, it is difficult to fill an aneurysm sac completely with coils, therefore, partial blocking of an aneurysm sac is inevitable. Blood flow characteristics, which may influence embolization process of an aneurysm, are affected by the locations of coils for partially blocked aneurysms. Blood flow fields inside an aneurysm are also influenced by the geometry of a parent vessel. In order to suggest the coil locations effective for aneurysm embolization, the blood flow fields of lateral aneurysm models were analyzed for different coil locations and parent vessel geometries. Flow rate into an aneurysm sac from a parent vessel (inflow rate) and wall shear stress were also calculated. Inflow rates were smaller and low wall shear regions were larger in the distal neck blocked model comparing to the dome blocked models. In the distal neck blocked model, inflow volume was smaller by 31% (straight parent vessel model) and 34% (curved parent vessel model) comparing to other models. The time averaged values of normalized low wall shear regions were 4% and 12% greater in the distal neck blocked models with a straight and a curved parent vessel, respectively. Since smaller inflow and low wall shear stress provide hemodynamic environment promoting thrombus embolization, distal neck should be the effective coil location for aneurysm embolization.

MeSH 主题词
Animals Blood Flow Velocity Blood Pressure Blood Vessel Prosthesis Carotid Arteries/physiopathology Computer Simulation Embolization, Therapeutic/methods Humans Intracranial Aneurysm/physiopathology,therapy Models, Cardiovascular Prosthesis Implantation/methods Shear Strength Treatment Outcome
作者与单位
共 2 位作者,点击展开单位 / ORCID
Byun Hong Sik
Department of Radiology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, South Korea.
Rhee Kyehan
Article Info
Journal
Medical engineering & physics
Abbr.
Med Eng Phys
ISSN
1350-4533
Published
2004-11-00
页码
755-61
Language
English
Country/Region
England
NLM ID
9422753
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]