Evaluation of high performance data acquisition boards for simultaneous sampling of fast signals from PET detectors.

Physics in medicine and biology ·Vol. 50 ·No. 1 ·2005-01-07 ·页码 29-44

Judenhofer MS, Pichler BJ, Cherry SR

Abstract

Detectors used for positron emission tomography (PET) provide fast, randomly distributed signals that need to be digitized for further processing. One possibility is to sample the signals at the peak initiated by a trigger from a constant fraction discriminator (CFD). For PET detectors, simultaneous acquisition of many channels is often important. To develop and evaluate novel PET detectors, a flexible, relatively low cost and high performance laboratory data acquisition (DAQ) system is therefore required. The use of dedicated DAQ systems, such as a multi-channel analysers (MCAs) or continuous sampling boards at high rates, is expensive. This work evaluates the suitability of well-priced peripheral component interconnect (PCI)-based 8-channel DAQ boards (PD2-MFS-8 2M/14 and PD2-MFS-8-500k/14, United Electronic Industries Inc., Canton, MA, USA) for signal acquisition from novel PET detectors. A software package was developed to access the board, measure basic board parameters, and to acquire, visualize, and analyse energy spectra and position profiles from block detectors. The performance tests showed that the boards input linearity is >99.2% and the standard deviation is <9 mV at 10 V for constant signals. Synchronous sampling of multiple channels and external synchronization of more boards are possible at rates up to 240 kHz per channel. Signals with rise times as fast as 130 ns (<2 V amplitude) can be acquired without slew rate effects. However, for signals with amplitudes of up to 5 V, a rise time slower than 250 ns is required. The measured energy resolution of a lutetium oxyorthosilicate (LSO)-photomultiplier tube (PMT) detector with a 22Na source was 14.9% (FWHM) at 511 keV and is slightly better than the result obtained with a high-end single channel MCA (8000A, Amptek, USA) using the same detector (16.8%). The crystals (1.2 x 1.2 x 12 mm3) within a 9 x 9 LSO block detector could be clearly separated in an acquired position profile. Thus, these boards are well suited for data acquisition with novel detectors developed for nuclear imaging.

MeSH 主题词
Computers Equipment Design Humans Image Processing, Computer-Assisted/methods Light Lutetium/chemistry Oscillometry Positron-Emission Tomography/instrumentation,methods Silicates/chemistry Software Time Factors
化学物质
Silicates Lutetium
作者与单位
共 3 位作者,点击展开单位 / ORCID
Judenhofer Martin S
Department of Biomedical Engineering, University of California, Davis, 95616, USA. [email protected]
Pichler Bernd J
Cherry Simon R
Article Info
Journal
Physics in medicine and biology
Abbr.
Phys Med Biol
ISSN
0031-9155
Corresponding email
Published
2005-01-07
页码
29-44
Language
English
Country/Region
England
NLM ID
0401220
基金资助
NIBIB NIH HHS · R01 EB00230 · United States
NIBIB NIH HHS · R01 EB00561 · United States
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]