A validated computational fluid dynamics model to estimate hemolysis in a rotary blood pump.

Artificial organs ·Vol. 29 ·No. 7 ·2005-07-00 ·页码 531-40

Arvand A, Hormes M, Reul H

Abstract

A major part of developing rotary blood pumps requires the optimization of hemolytic properties of the entire pump. Application of a suited computational fluid dynamics (CFD)-based hemolysis model allows approximation of blood damage in an early phase of the design process. Thus, a drastic reduction of time- and cost- intensive hemolysis experiments can be achieved. For the MicroDiagonal Pump (MDP), still under development at Helmholtz-Institute in Aachen, Germany, different pump configurations have been analyzed, both numerically and experimentally. The CFD model of the pump has been successfully validated based on the comparison of the pressure head curves (H-Q curves), as discussed in a prior publication. In the present study, the authors focus on the development of a semiempiric blood damage model using the CFD and in vitro hemolysis data. On the one hand, mean key characteristics (shear stress and exposure time) and other characteristics affecting blood damage have been calculated based on numerical data. On the other hand, in vitro hemolysis tests have been accomplished in order to determine the hemolytic curves of two different pump configurations (with the same impeller but different tip clearances). Finally, a new function based on a general power law has been defined by means of the mean key characteristics. The unknown constants of the function have been determined by multidimensional regression analysis using the hemolytic curves. For the final validation of this new blood damage model, the calculated and the in vitro obtained hemolysis indices at the specific VAD operating point have been compared for all pump configurations. The comparison showed an excellent agreement, both qualitatively and quantitatively.

MeSH 主题词
Animals Computational Biology Equipment Design Equipment Failure Analysis Heart-Assist Devices Hemolysis Hemorheology Humans In Vitro Techniques Models, Cardiovascular Reproducibility of Results Rotation Swine
作者与单位
共 3 位作者,点击展开单位 / ORCID
Arvand Arash
Helmholtz-Institute for Biomedical Engineering, RWTH Aachen, Aachen, Germany. [email protected]
Hormes Marcus
Reul Helmut
Article Info
Journal
Artificial organs
Abbr.
Artif Organs
ISSN
0160-564X
Corresponding email
Published
2005-07-00
页码
531-40
Language
English
Country/Region
United States
NLM ID
7802778
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]