Development of high throughput optical sensor array for on-line pH monitoring in micro-scale cell culture environment.

Biomedical microdevices ·Vol. 11 ·No. 1 ·2009-02-00 ·页码 265-73

Wu MH, Lin JL, Wang J, Cui Z, Cui Z

Abstract

On-line pH detection of cell culture environment is necessary in a bioprocess or tissue engineering. Devices by means of electrochemical mechanisms for this purpose have been reported to be less suitable compared with optical-based sensing principles. More recently, some non-invasive optical sensing systems have been proposed for online pH monitoring of cell culture environment. However, these devices are not for multi-target pH monitoring purpose, and are large in scale and thus not appropriate for the pH monitoring at a micro scale such as in microbioreactor or microfluidic-based cell culture platform. To tackle these issues, an optical fiber sensor array for on-line pH monitoring was proposed using microfluidic technology. The working principle is based on the optical absorption of phenol red normally contained in culture medium. Different from other device of the similar working principle, the proposed device requires less liquid volume (less than 0.8 microl), is non-invasive, and particularly can be configured as an array for high throughput pH monitoring. The present device has been optimized for the shape of detection chamber in a microfluidic chip with the aid of computational fluid dynamics (CFD) simulation, to avoid flow dead zone and thus to reduce the response time of detection. Both simulation and experimental results revealed that the design of oval detection chamber (axis, 1.5 and 2.0 mm) can considerably reduce the response time. Preliminary test has proved that the optical pH detection device is able to detect pH with average detection sensitivity of 0.83 V/pH in the pH range of 6.8-7.8, which is normally experienced in mammalian cell culture.

MeSH 主题词
Animals Cell Culture Techniques/instrumentation,methods Computer Simulation Electrochemical Techniques/instrumentation,methods Humans Hydrogen-Ion Concentration Microfluidic Analytical Techniques/instrumentation,methods Optics and Photonics/instrumentation,methods Sensitivity and Specificity
作者与单位
共 5 位作者,点击展开单位 / ORCID
Wu Min-Hsien
Graduate Institute of Biochemical and Biomedical Engineering, Chang Gung University, Taoyuan, Taiwan.
Lin Jr-Lung
Wang Junbo
Cui Zhanfeng
Cui Zheng
Article Info
Journal
Biomedical microdevices
Abbr.
Biomed Microdevices
ISSN
1572-8781
Published
2009-02-00
页码
265-73
Language
English
Country/Region
United States
NLM ID
100887374
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]