Precision and stability of dual-energy X-ray absorptiometry measurements.

Calcified tissue international ·Vol. 49 ·No. 3 ·1991-09-00 ·Pages 174-8

Johnson J, Dawson-Hughes B

Abstract

This study was performed to determine the precision and stability of dual-energy X-ray absorptiometry (DEXA) measurements, to compare bone mineral density (BMD) of subjects measured by DEXA and radionuclide dual-photon absorptiometry (DPA), and to evaluate different absorber materials for use with an external standard. Short-term precision (% coefficient of variation, CV) was determined in 6 subjects scanned six times each with repositioning, initially and 9 months later. Mean CV was 1.04% for spine and 2.13% for femoral neck BMD; for whole-body measurements in 5 subjects, mean CV was 0.64% for BMD, 2.2% for fat, and 1.05% for lean body mass. Precision of aluminum phantom measurements made over a 9-month period was 0.89% with the phantom in 15.2 cm, 0.88% in 20.3 cm, and 1.42% in 27.9 cm of water. In 51 subjects, BMD by DEXA and DPA was correlated for the spine (r = 0.98, P = 0.000) and femoral neck (r = 0.91, P = 0.000). Spine BMD was 4.5% lower and femoral neck BMD 3.1% higher by DEXA than by DPA. An aluminum phantom was scanned repeatedly, in both water and in an oil/water (30:70) mixture at thickness ranging from 15.2 through 27.9 cm. Phantom BMD was lower at 15.2 cm than at higher thicknesses of both water and oil/water (P = 0.05, ANOVA). The phantom was scanned repeatedly in 15.2, 20.3, and 27.9 cm of water over a 9 month period. In 15.2 and 20.3 cm of water, phantom BMD did not vary significantly whereas in 27.9 cm of water (equivalent to a human over 30 cm thick), phantom BMD increased 2.3% (P = 0.01) over the 9 months.

MeSH Terms
Absorptiometry, Photon/standards Aged Bone Density Female Femur Humans Middle Aged Spine
Authors & Affiliations
2 authors, click to expand affiliations / ORCID
Johnson J
Calcium and Bone Metabolism Lab, USDA Human Nutrition Research Center on Aging, Tufts University, Boston, MA 02111.
Dawson-Hughes B
References (12)
12 references, click to expand
  1. A comparison of quantitative dual-energy radiographic absorptiometry and dual photon absorptiometry of the lumbar spine in postmenopausal women.
    Calcif Tissue Int. 1989 Nov;45(5):288-91 PMID: 2509017
  2. Comparison of dual-energy x-ray absorptiometry and dual photon absorptiometry for bone mineral measurements of the lumbar spine.
    Mayo Clin Proc. 1988 Nov;63(11):1075-84 PMID: 3193817
  3. Correction of the effects of source, source strength, and soft-tissue thickness on spine dual-photon absorptiometry measurements.
    Calcif Tissue Int. 1989 Apr;44(4):251-7 PMID: 2501007
  4. Dual-photon absorptiometry: comparison of bone mineral and soft tissue mass measurements in vivo with established methods.
    Am J Clin Nutr. 1989 Jun;49(6):1283-9 PMID: 2729167
  5. Precision error in dual-photon absorptiometry related to source age.
    Radiology. 1988 Feb;166(2):523-7 PMID: 3336729
  6. Quantitative digital radiography versus dual photon absorptiometry of the lumbar spine.
    J Clin Endocrinol Metab. 1988 Oct;67(4):839-44 PMID: 3417851
  7. Dual energy radiography (DER): a preliminary comparative study.
    Calcif Tissue Int. 1988 Sep;43(3):189-91 PMID: 3141022
  8. Measurement of bone by dual-photon absorptiometry (DPA) and dual-energy X-ray absorptiometry (DEXA).
    Ann Chir Gynaecol. 1988;77(5-6):197-203 PMID: 3076046
  9. X-ray dual-photon absorptiometry: a new method for the measurement of bone density.
    Br J Radiol. 1989 Jul;62(739):587-92 PMID: 2758245
  10. Long-term precision of dual photon absorptiometry in the lumbar spine in clinical settings.
    Bone Miner. 1988 Mar;3(4):305-15 PMID: 3228615
  11. Dual photon absorptiometric measurements of bone mineral density increase with source life.
    Calcif Tissue Int. 1987 Nov;41(5):293-4 PMID: 3121152
  12. Performance evaluation of a dual-energy x-ray bone densitometer.
    Calcif Tissue Int. 1989 Mar;44(3):228-32 PMID: 2493329
Article Info
Journal
Calcified tissue international
Abbr.
Calcif Tissue Int
ISSN
0171-967X
Published
1991-09-00
Pages
174-8
Language
English
Region
United States
NLM ID
7905481
Subset
IM
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]