Abstract
An estimated 10,000-15,000 pacemaker and implantable cardioverter-defibrillator (ICD) leads are extracted annually worldwide using specialized tools that disrupt encapsulating fibrous tissue. Additional information is needed regarding the safety of the devices that have been approved for lead extraction. The aim of this study was to determine whether complications due to device-assisted lead extraction might be more hazardous than published data suggest, and whether procedural safety precautions are effective. We searched the US Food and Drug Administration's (FDA) Manufacturers and User Defined Experience (MAUDE) database from 1995 to 2008 using the search terms 'lead extraction and death' and 'lead extraction and injury'. Additional product specific searches were performed for the terms 'death' and 'injury'. Between 1995 and 2008, 57 deaths and 48 serious cardiovascular injuries associated with device-assisted lead extraction were reported to the FDA. Owing to underreporting, the FDA database does not contain all adverse events that occurred during this period. Of the 105 events, 27 deaths and 13 injuries occurred in 2007-2008. During these 2 years, 23 deaths were linked with excimer laser or mechanical dilator sheath extractions. The majority of deaths and injuries involved ICD leads, and most were caused by lacerations of the right atrium, superior vena cava, or innominate vein. Overall, 62 patients underwent emergency surgical repair of myocardial perforations and venous lacerations and 35 (56%) survived. These findings suggest that device-assisted lead extraction is a high-risk procedure and that serious complications including death may not be mitigated by emergency surgery. However, skilled standby cardiothoracic surgery is essential when performing pacemaker and ICD lead extractions. Although the incidence of these complications is unknown, the results of our study imply that device-assisted lead extractions should be performed by highly qualified physicians and their teams in specialized centres.
MeSH Terms
Arrhythmias, Cardiac/mortality,therapy
Databases, Factual
Death, Sudden, Cardiac/epidemiology,etiology
Defibrillators, Implantable/adverse effects,statistics & numerical data
Electrodes, Implanted/adverse effects,statistics & numerical data
Equipment Failure Analysis/statistics & numerical data
Equipment Safety/statistics & numerical data
Heart Injuries/etiology,mortality
Humans
Pacemaker, Artificial/adverse effects,statistics & numerical data
Product Surveillance, Postmarketing
Prosthesis Implantation/mortality
Risk Factors
United States/epidemiology
United States Food and Drug Administration/statistics & numerical data
Authors & Affiliations
5 authors, click to expand affiliations / ORCID
Hauser Robert G
Minneapolis Heart Institute Foundation, 920 East 28th Street, Suite 300, Minneapolis, MN 55407, USA.
[email protected]
Katsiyiannis William T
Gornick Charles C
Almquist Adrian K
Kallinen Linda M
References (21)
21 references, click to expand
-
Task Force 6: training in specialized electrophysiology, cardiac pacing, and arrhythmia management: endorsed by the Heart Rhythm Society.
J Am Coll Cardiol. 2006 Feb 21;47(4):904-10
PMID: 16487872
-
Early failure of a small-diameter high-voltage implantable cardioverter-defibrillator lead.
Heart Rhythm. 2007 Jul;4(7):892-6
PMID: 17599673
-
Laser-assisted lead extraction: the European experience.
Europace. 2007 Aug;9(8):651-6
PMID: 17597078
-
Lead extraction: a new effective tool to overcome fibrous binding sites.
J Interv Card Electrophysiol. 2009 Mar;24(2):147-50
PMID: 18987963
-
Large, single-center, single-operator experience with transvenous lead extraction: outcomes and changing indications.
Heart Rhythm. 2008 Apr;5(4):520-5
PMID: 18325849
-
Learning curve characteristics of pacing lead extraction with a laser sheath.
Pacing Clin Electrophysiol. 1998 Nov;21(11 Pt 2):2309-13
PMID: 9825338
-
Intravascular extraction of problematic or infected permanent pacemaker leads: 1994-1996. U.S. Extraction Database, MED Institute.
Pacing Clin Electrophysiol. 1999 Sep;22(9):1348-57
PMID: 10527016
-
Pacemaker lead extraction with the laser sheath: results of the pacing lead extraction with the excimer sheath (PLEXES) trial.
J Am Coll Cardiol. 1999 May;33(6):1671-6
PMID: 10334441
-
Unexpected ICD pulse generator failure due to electronic circuit damage caused by electrical overstress.
Pacing Clin Electrophysiol. 2001 Jul;24(7):1046-54
PMID: 11475818
-
Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management: this document was endorsed by the American Heart Association (AHA).
Heart Rhythm. 2009 Jul;6(7):1085-104
PMID: 19560098
-
Clinical study of the laser sheath for lead extraction: the total experience in the United States.
Pacing Clin Electrophysiol. 2002 May;25(5):804-8
PMID: 12049372
-
Lead extraction.
Heart Rhythm. 2007 Sep;4(9):1238-43
PMID: 17765629
-
Five-years experience with intravascular lead extraction. U.S. Lead Extraction Database.
Pacing Clin Electrophysiol. 1994 Nov;17(11 Pt 2):2016-20
PMID: 7845810
-
Nonthoracotomy implantable defibrillator lead extraction: results and comparison with extraction of pacemaker leads.
Pacing Clin Electrophysiol. 2003 Oct;26(10):1944-50
PMID: 14516333
-
Deaths associated with implantable cardioverter defibrillator failure and deactivation reported in the United States Food and Drug Administration Manufacturer and User Facility Device Experience Database.
Heart Rhythm. 2004 Oct;1(4):399-405
PMID: 15851191
-
Pacemaker and ICD lead extraction with electrosurgical dissection sheaths and standard transvenous extraction systems: results of a randomized trial.
Europace. 2007 Feb;9(2):98-104
PMID: 17272329
-
Increasing hazard of Sprint Fidelis implantable cardioverter-defibrillator lead failure.
Heart Rhythm. 2009 May;6(5):605-10
PMID: 19285920
-
Is there an adverse outcome from abandoned pacing leads?
J Interv Card Electrophysiol. 2000 Oct;4(3):493-9
PMID: 11046188
-
Intravascular lead extraction using the excimer laser: pitfalls and tips for success.
Semin Vasc Surg. 2008 Mar;21(1):54-6
PMID: 18342737
-
Laser lead extraction: predictors of success and complications.
Pacing Clin Electrophysiol. 2007 Feb;30(2):214-20
PMID: 17338718
-
How to prevent, recognize, and manage complications of lead extraction. Part III: Procedural factors.
Heart Rhythm. 2008 Sep;5(9):1352-4
PMID: 18614405