Journal of veterinary internal medicine ·Vol. 24 ·No. 1 ·2010-00-00 ·Pages 120-6

Moretti S, Tschuor F, Osto M, Franchini M, Wichert B, Ackermann M, Lutz TA, Reusch CE, Zini E

Abstract

The Guardian REAL-Time is a continuous glucose-monitoring system (CGMS) recently developed to provide instantaneous interstitial glucose concentrations; the system does not require a monitor being fixed to the animal. The CGMS provides accurate and reproducible real-time readings of glucose concentration in cats. Thirty-two diabetic cats, 2 cats with suspected insulinoma, and 5 healthy cats. Prospective, observational study. CGMS accuracy was compared with a reference glucose meter at normal, high, and low blood glucose concentrations using error grid analysis. Reading variability of 2 simultaneously used CGMS was determined in diabetic cats by calculating correlation and percentage of concordance of paired data at different glycemic ranges. The time interval between increasing glycemia and a rise in interstitial fluid glucose measured by the CGMS was assessed in healthy cats receiving glucose IV; the time point of maximal increase in interstitial glucose concentrations was calculated. The CGMS was 100, 96.1, and 91.0% accurate at normal, high, and low blood glucose concentrations. Measurements deviated from reference by -12.7 +/- 70.5 mg/dL at normal, -12.1 +/- 141.5 mg/dL at high, and -1.9 +/- 40.9 mg/dL at low glucose concentrations. Overall, paired CGMS readings correlated significantly (r = 0.95, P < .0001) and concordance was 95.7%. The median delay after IV administration of glucose to an increase in interstitial glucose was 11.4 minutes (range: 8.8-19.7 minutes). Although some readings substantially deviated from reference values, the CGMS yields reproducible results, is clinically accurate in cats with hyperglycemia and euglycemia, and is slightly less accurate if blood glucose concentrations are low. Rapidly increasing interstitial glucose after a glycemic rise suggests that the CGMS is suitable for real-time measurement under clinical conditions.

MeSH Terms
Animals Blood Glucose/analysis Cat Diseases/blood Cats Diabetes Mellitus/blood,veterinary Female Male Monitoring, Physiologic/instrumentation,veterinary Sensitivity and Specificity
Chemicals
Blood Glucose
Authors & Affiliations
9 authors, click to expand affiliations / ORCID
Moretti S
Clinic for Small Animal Internal Medicine, Vetsuisse Faculty, University of Zürich, Winterthurerstr 260, 8057 Zürich, Switzerland.
Tschuor F
Osto M
Franchini M
Wichert B
Ackermann M
Lutz T A
Reusch C E
Zini E
Article Info
Journal
Journal of veterinary internal medicine
Abbr.
J Vet Intern Med
ISSN
0891-6640
Published
2010-00-00
Pages
120-6
Language
English
Region
United States
NLM ID
8708660
Subset
IM
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]