Home LiteratureArticle Details
PMID: 20376278 Published · ppublish English Journal Article

Potentials of Digitally Sampling Scintillation Pulses in Timing Determination in PET.

IEEE transactions on nuclear science ·Vol. 56 ·No. 5 ·2009-10-06 ·页码 2607-2613

Xie Q, Kao CM, Wang X, Guo N, Zhu C, Frisch H, Moses WW, Chen CT

Abstract

We investigate the potentials of digitally sampling scintillation pulses techniques for positron emission tomography (PET) in this paper, focusing on the determination of the event time. We have built, and continue building, a digital library of PET event waveforms generated with various combinations of photo-detectors and scintillator materials, with various crystal sizes. Events in this digital library are obtained at a high sampling of 20 GSps (Giga-samples per second) so that their waveforms are recorded with high accuracy. To explore the potential advantages of digitally sampling scintillation pulses, we employ a dataset in the above-mentioned library to evaluate two methods for digitizing the event pulses and linear interpolation techniques to analyze the resulting digital samples. Our results show that the two digitization methods that we studied can yield a coincidence timing resolution of about 300 ps FWHM when applied to events generated by a pair of LSO + PMT detector units. This timing resolution is comparable with that is achieved by the same detector pair with a constant fraction discriminator (CFD). As a benchmark, regular-time sampling (RTS) method, usually implemented with very fast traditional analog-to-digital converters (ADCs) for digitizing scintillation pulses, is not feasible for a multi-channel system like a PET system. Digitizing scintillation pulses with multi-voltage threshold (MVT) method could be implemented at a reasonable cost for a PET system. With digitized PET event samples, various digital signal processing (DSP) techniques can be implemented to determine event arrival time. Our results have therefore demonstrated the promising potentials of digitally sampling scintillation pulses techniques in PET imaging.

作者与单位
共 8 位作者,点击展开单位 / ORCID
Xie Qingguo
Wuhan National Laboratory for Optoelectronics, Wuhan, Hubei 430074, China and Biomedical Engineering Department, Huazhong University of Science and Technology, Wuhan, Hubei 430074, China.
Kao Chien-Min
Wang Xi
Guo Ning
Zhu Caigang
Frisch Henry
Moses William W
Chen Chin-Tu
Article Info
Journal
IEEE transactions on nuclear science
Abbr.
IEEE Trans Nucl Sci
ISSN
0018-9499
Published
2009-10-06
页码
2607-2613
Language
English
Country/Region
United States
NLM ID
9879488
基金资助
NIBIB NIH HHS · R01 EB006085 · United States
NIBIB NIH HHS · R01 EB006085-03 · United States
NIBIB NIH HHS · R01 EB000225 · United States
NIBIB NIH HHS · R33 EB001928 · United States
NIBIB NIH HHS · R01 EB006085-02 · United States
NIBIB NIH HHS · R01 EB006085-04 · United States
NIBIB NIH HHS · R33 EB001928-04 · United States
NIBIB NIH HHS · R33 EB001928-03 · United States
NIBIB NIH HHS · R21 EB001928-01 · United States
NIBIB NIH HHS · R01 EB006085-01A2 · United States
NIBIB NIH HHS · R33 EB001928-02 · United States
NIBIB NIH HHS · R21 EB001928 · United States
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]