Home LiteratureArticle Details
PMID: 21654494 Published · ppublish English Journal Article

Numerical and experimental flow analysis of the Wang-Zwische double-lumen cannula.

ASAIO journal (American Society for Artificial Internal Organs : 1992) ·Vol. 57 ·No. 4 ·2011-00-00 ·页码 318-27

De Bartolo C, Nigro A, Fragomeni G, Colacino FM, Wang D, Jones CC, Zwischenberger J

Abstract

An experimental and numerical analysis was performed for the Wang-Zwische double-lumen cannula (DLC) (Avalon Elite). The aim of this work was to provide insight for future improvement by characterizing the fluid dynamic behavior of the novel catheter with metrics often associated with blood trauma. Pressure and flow distributions were measured on a steady-flow rig using a 50% glycerol-water mixture by imposing a 2 L/min flow rate across the drainage and infusion lumens. The fluid was modeled as Newtonian with density of 1050 kg/m³ and dynamic viscosity of 0.0035 kg/m·s. Reynolds numbers typical for transitional flow (Re = 2000-2500) were computed within the lumens because of the changing cross-sections of the cannula geometry. Numerical computations were performed using the steady three-dimensional Reynolds-averaged Navier-Stokes (RANS) equations and the low-Reynolds k-ω turbulence model. Discretization of governing equations was based on a cell-centered finite volume method. Numerical results correlated well with global performance of the cannula, allowing evaluation of the geometry toward potential blood trauma. Peak wall shear stress (WSS) in the drainage lumen was higher than that of infusion lumen, mainly due to the presence of side holes. Furthermore, recirculation regions were predicted in transition tubing to connectors of both the drainage and the infusion lumens because of adverse pressure gradients caused by the sudden enlargement of the cannula geometry. In this three-dimensional computational fluid dynamics (CFD) study, we observed higher peak WSS values for the drainage lumen, which may potentially cause blood trauma. Furthermore, recirculation regions were predicted in the proximity of the exit sections of both the infusion and drainage lumens, which may contribute to thrombosis formation. This study provides insight for future DLC modifications in minimizing cannula-induced blood trauma and thrombogenicity in long-term applications.

MeSH 主题词
Blood Flow Velocity Cardiovascular Surgical Procedures/instrumentation Catheters Computer Simulation Equipment Design Extracorporeal Membrane Oxygenation Glycerol/chemistry Humans Hydrodynamics Models, Cardiovascular Pressure Stress, Mechanical Thrombosis/prevention & control Viscosity Water/chemistry
化学物质
Water Glycerol
作者与单位
共 7 位作者,点击展开单位 / ORCID
De Bartolo Carmine
Department of Mechanical Engineering, University of Calabria, Rende, Italy.
Nigro Alessandra
Fragomeni Gionata
Colacino Francesco M
Wang Dongfang
Jones Cameron C
Zwischenberger Joseph
Article Info
Journal
ASAIO journal (American Society for Artificial Internal Organs : 1992)
Abbr.
ASAIO J
ISSN
1538-943X
Published
2011-00-00
页码
318-27
Language
English
Country/Region
United States
NLM ID
9204109
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]