主页 文献库文献详情
PMID: 2302775 已发表 · ppublish 英语

Improved kinetic rate assay of urinary N-acetyl-beta-D-glucosaminidase with 2-chloro-4-nitrophenyl-N-acetyl-beta-D-glucosaminide as substrate.

Clinical chemistry ·第 36 卷 ·第 2 期 ·1990-03-23

Makise J, Saito E, Obuchi M, Kanayama M, Harakawa K, Yoshida K

摘要

We have improved the kinetic rate assay method for determining N-acetyl-beta-D-glucosaminidase (EC 3.2.1.30; NAG) activity in urine with use of the synthetic substrate, 2-chloro-4-nitrophenyl-N-acetyl-beta-D-glucosaminide (CNP-NAG), reported previously (Clin Chem 1988;34:2140-2). To increase the solubility of this substrate, we used crown ether (15-crown-5-ether) and ethylene glycol. In addition, we used for the standard solution NAG from human placenta, with specificity corresponding to that of human urine, so that values obtained with the CNP-NAG method and a p-nitrophenyl-NAG method ("MEI Assay NAG") correlated almost completely (r = 0.995, n = 29). Reference values for urinary NAG activity determined by the CNP-NAG method were established for untimed urine specimens from 674 healthy volunteers. The normal reference interval (mean +/- 2 SD) for NAG: 1.6-15.0 (mean 4.9) U per gram of creatinine.

文献信息
期刊
Clinical chemistry
期刊简称
Clin Chem
发表日期
1990-03-23
收录日期
1990-03-23
更新日期
2013-11-21
语言
英语
国家/地区
United States
NLM ID
9421549
分析服务
分析服务

联系地址

山东省济南市章丘区文博路2号

齐鲁师范学院 genelibs生信实验室

山东省济南市高新区舜华路750号

大学科技园北区F座4单元2楼

电话: 0531-88819269

微信公众号

关注微信订阅号,实时查看信息,关注医学生物学动态。


商务邮箱

E-mail: [email protected]