Computational fluid dynamic analysis to prevent aortic root and valve clots during left ventricular assist device support.

The journal of extra-corporeal technology ·Vol. 44 ·No. 4 ·2012-12-00 ·页码 210-5

Poullis M

Abstract

Aortic root and valve clots are rare but well described in patients on maximal left ventricular assist device (LVAD) support. We performed a theoretical analysis using computational fluid dynamic analyses in two dimensions to try and ascertain if inflow cannula design/orientation/placement affect aortic root flow dynamics. Two-dimensional computational fluid dynamics using easy CFD-G was performed. The effect of a curved inflow cannula, a straight cannula, and one with a hole in the outer curve was analyzed. In addition, the effect of inflow conduit angulation on the ascending aorta was studied. Computational fluid dynamic (CFD) analysis predicts that stagnant blood exists in the aortic root when little or no cardiac ejection is taking place. Coronary flow is too small to affect the root flow streamlines. A hole on the root side of a curved inflow aortic cannula increases the flow in the aortic root and may decrease the incidence of root and valve thrombosis. The angle of the inflow conduit attachment to the ascending aorta was also found to be crucial with regard to aortic root blood stasis. In addition, a baffle at the tip of the inflow cannula may prove to be beneficial. Theoretical analysis using the technique of CFD predicts that inflow cannula position and design may affect the incidence of aortic root thrombosis during LVAD support when minimal cardiac ejection is occurring.

MeSH 主题词
Aorta/anatomy & histology,physiology Computer Simulation Heart Ventricles/anatomy & histology Heart-Assist Devices Hemodynamics Humans Models, Cardiovascular Thrombosis/prevention & control Ventricular Function/physiology
作者与单位
共 1 位作者,点击展开单位 / ORCID
Poullis Michael
Liverpool Heart and Chest Hospital, Liverpool, United Kingdom. [email protected]
Article Info
Journal
The journal of extra-corporeal technology
Abbr.
J Extra Corpor Technol
ISSN
0022-1058
Corresponding email
Published
2012-12-00
页码
210-5
Language
English
Country/Region
United States
NLM ID
0267637
External Links
PubMed source
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]