Postimplant device thrombosis remains a life-threatening complication and limitation of continuous-flow ventricular assist devices (VADs). Using advanced computational fluid dynamic (CFD) simulations, we successfully depicted various flow patterns, recirculation zones, and stagnant platelet trajectories which promote thrombus formation and observed that they matched actual thrombus formation patterns observed in Thoratec HeartMate II VADs explanted from patients with pump thrombosis. Previously, these small eddies could not be captured by either digital particle image velocimetry or CFD due to insufficient resolution. Our study successfully demonstrated the potential capability of advanced CFD to be adopted for device optimization, leading to enhanced safety and efficacy of VADs for long-term destination therapy.