Influence of ventricular fibrillation duration on defibrillation energy in dogs using bidirectional pulse discharges.

Pacing and clinical electrophysiology : PACE ·Vol. 11 ·No. 9 ·1988-09-00 ·Pages 1315-23

Echt DS, Barbey JT, Black JN

Abstract

The automatic implantable defibrillator device typically discharges 5-30 seconds after detection of ventricular fibrillation. To investigate the importance of the duration of ventricular fibrillation on defibrillation, the effects of ventricular fibrillation durations of 5, 15, and 30 seconds on the energy requirements for successful internal defibrillation were compared in 15 closed chest dogs with internal electrodes. The electrode configuration utilized a transvenous right heart catheter with two electrodes and a precordial subcutaneous patch electrode, with a single bidirectional pulse discharged between the distal catheter electrode and the proximal catheter and patch electrodes. Curves of energy vs. percentage of successful defibrillation were constructed and logistic regression was used to derive 90% and 50% successful energy doses (ED90 and ED50). The mean ventricular fibrillation activation interval just prior to defibrillation was determined from discrete RV endocardial electrograms. Four dogs died during testing, all because of inability to defibrillate after 30 s of ventricular fibrillation. In the remaining 11 dogs, the ED90 increased from (mean +/- SD) 27 +/- 13J at 5 s to 41 +/- 14J at 30 s (p less than .01). The mean ventricular fibrillation activation interval decreased from 107 +/- 21 ms at 5 s to 95 +/- 18 ms at 30 s (p less than .01). In conclusion, the energy required for internal defibrillation in dogs using this electrode configuration increases with longer durations of ventricular fibrillation, and is associated with more rapid ventricular fibrillation activation intervals.

MeSH Terms
Animals Blood Pressure Cardiac Catheterization/instrumentation Dogs Electric Countershock/instrumentation,methods Electrocardiography Electrodes, Implanted Equipment Design Pacemaker, Artificial Prostheses and Implants Random Allocation Time Factors Ventricular Fibrillation/physiopathology
Authors & Affiliations
3 authors, click to expand affiliations / ORCID
Echt D S
Department of Medicine, Vanderbilt University, School of Medicine, Nashville, Tennessee 37232.
Barbey J T
Black J N
Article Info
Journal
Pacing and clinical electrophysiology : PACE
Abbr.
Pacing Clin Electrophysiol
ISSN
0147-8389
Published
1988-09-00
Pages
1315-23
Language
English
Region
United States
NLM ID
7803944
Subset
IM
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]