The study on hemodynamic effect of varied support models of BJUT-II VAD on coronary artery: a primary CFD study.

ASAIO journal (American Society for Artificial Internal Organs : 1992) ·Vol. 60 ·No. 6 ·2014-00-00 ·页码 643-51

Zhang Q, Gao B, Gu K, Chang Y, Xu J

Abstract

BJUT-II VAD (Beijing University of Technology ventricular assist device II) is a novel left ventricular assist device. Because of the special connection between the pump and native heart, the hemodynamic effects of BJUT-II VAD on coronary artery are still unclear. Hence, numerical simulations have been conducted to clarify changes in hemodynamic effects of different support modes. A patient-specific left coronary arterial geometric model is reconstructed based on the computed tomography (CT) data. Three support modes, "constant speed mode," "co-pulse mode," and "counter pulse mode," are used in this study. The wall shear stress (WSS), wall shear stress gradient (WSSG), cycle-averaged wall shear stress (avWSS), oscillatory shear index (OSI), and the flow pattern are calculated to evaluate the hemodynamic states of coronary artery. The computational results demonstrate that the hemodynamic states of coronary artery are directly affected by the support modes. The co-pulse modes could achieve the highest blood perfusion (constant speed: 153 ml/min vs. co-pulse: 775 ml/min vs. counter pulse: 140 ml/min) and the highest avWSS (constant speed: 18.1 Pa vs. co-pulse: 42.6 Pa vs. counter pulse: 22.6 Pa). In addition, both the WSS and WSSG at the time of peak blood velocity under the constant speed mode are lower than those under other two support modes. In contrast, the counter pulse mode generates the highest OSI value (constant speed: 0.365 vs. co-pulse: 0.379 vs. counter pulse: 0.426). BJUT-II VAD under co-pulse mode may have benefits for improving coronary perfusion and preventing the development of atherosclerosis; however, the constant speed mode may have benefit for preventing the development of plaque vulnerability.

MeSH 主题词
Biomechanical Phenomena Computer Simulation Coronary Circulation Coronary Vessels/anatomy & histology,physiology,surgery Heart Failure/physiopathology,surgery Heart-Assist Devices Hemodynamics Humans Hydrodynamics Models, Cardiovascular
作者与单位
共 5 位作者,点击展开单位 / ORCID
Zhang Qi
From the School of Life Science and BioEngineering, Beijing University of Technology, Beijing, P.R. China.
Gao Bin
Gu Kaiyun
Chang Yu
Xu Jinchao
Article Info
Journal
ASAIO journal (American Society for Artificial Internal Organs : 1992)
Abbr.
ASAIO J
ISSN
1538-943X
Published
2014-00-00
页码
643-51
Language
English
Country/Region
United States
NLM ID
9204109
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]