Comparison Between Bench-Top and Computational Modelling of Cerebral Thromboembolism in Ventricular Assist Device Circulation.

Cardiovascular engineering and technology ·Vol. 6 ·No. 3 ·2015-09-00 ·页码 242-55

Clark WD, Eslahpazir BA, Argueta-Morales IR, Kassab AJ, Divo EA, DeCampli WM

Abstract

Despite improvements in ventricular assist devices (VAD) design, VAD-induced stroke rates remain remarkably high at 14-47%. We previously employed computational fluid dynamics (CFD) to propose adjustment of VAD outflow graft (VAD-OG) implantation to reduce stoke. Herein, we present an in-vitro model of cerebral vessel embolization in VAD-assisted circulation, and compare benchtop results to CFD predictions. The benchtop flow-loop consists of a 3D printed aortic bed using Accura 60 polymer driven by a continuous-flow pump. Three hundred spherical particles simulating thrombi of 2, 3.5, and 5 mm diameters were injected at the mock VAD-OG inlet. A water and glycerin mixture (3.8 cP viscosity) synthetically mimicked blood. The flowrate was adjusted to match the CFD Reynolds number. Catch cans were used to capture and count particles reaching cerebral vessels. VAD-OG geometries were evaluated using comparison of means Z-score range of -1.96 ≤ Z ≤ 1.96 to demonstrate overall agreement between computational and in-vitro techniques. Z-scores were: (i) Z = -1.05 for perpendicular (0°), (ii) Z = 0.32 for intermediate (30°), and (iii) Z = -0.52 for shallow (60°) anastomosis and confirmed agreement for all geometries. This study confirmed added benefits of using a left carotid artery bypass-graft with percent embolization reduction: 22.6% for perpendicular, 21.2% for intermediate, and 11.9% for shallow anastomoses. The shallow anastomosis demonstrated lower degrees of aortic arch flow recirculation, consistent with steady-flow computations. Quantitatively and qualitatively, contemporary steady-flow computational models for predicting VAD-induced cerebral embolization can be achieved in-vitro to validate the CFD equivalent.

Keywords
Computational fluid dynamics In-vitro cardiovascular modeling Left ventricular assist device Stroke
MeSH 主题词
Carotid Arteries/physiopathology Cerebral Cortex/blood supply Coronary Artery Bypass/methods Heart-Assist Devices Hydrodynamics Intracranial Thrombosis/physiopathology Models, Cardiovascular
作者与单位
共 6 位作者,点击展开单位 / ORCID
Clark William D
Department of Mechanical and Aerospace Engineering, University of Central Florida, 4000 Central Florida Blvd., Orlando, FL, 32816, USA.
Eslahpazir Benjamin A
College of Medicine, University of Central Florida, Orlando, FL, USA.
Argueta-Morales I Ricardo
Division of Cardiothoracic Surgery, The Heart Center at Arnold Palmer Hospital for Children, Orlando, FL, USA.
Kassab Alain J
Department of Mechanical and Aerospace Engineering, University of Central Florida, 4000 Central Florida Blvd., Orlando, FL, 32816, USA. [email protected].
Divo Eduardo A
Department of Mechanical Engineering, Embry-Riddle Aeronautical University, Daytona Beach, FL, USA.
DeCampli William M
College of Medicine, University of Central Florida, Orlando, FL, USA. | Division of Cardiothoracic Surgery, The Heart Center at Arnold Palmer Hospital for Children, Orlando, FL, USA.
Article Info
Journal
Cardiovascular engineering and technology
Abbr.
Cardiovasc Eng Technol
ISSN
1869-4098
Corresponding email
Published
2015-09-00
电子出版
2015-00-23
页码
242-55
Language
English
Country/Region
United States
NLM ID
101531846
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]