Home LiteratureArticle Details
PMID: 28114192 Published · ppublish English Journal Article Validation Study

FDA Benchmark Medical Device Flow Models for CFD Validation.

ASAIO journal (American Society for Artificial Internal Organs : 1992) ·Vol. 63 ·No. 2 ·2017-00-00 ·页码 150-160

Malinauskas RA, Hariharan P, Day SW, Herbertson LH, Buesen M, Steinseifer U, Aycock KI, Good BC, Deutsch S, Manning KB, Craven BA

Abstract

Computational fluid dynamics (CFD) is increasingly being used to develop blood-contacting medical devices. However, the lack of standardized methods for validating CFD simulations and blood damage predictions limits its use in the safety evaluation of devices. Through a U.S. Food and Drug Administration (FDA) initiative, two benchmark models of typical device flow geometries (nozzle and centrifugal blood pump) were tested in multiple laboratories to provide experimental velocities, pressures, and hemolysis data to support CFD validation. In addition, computational simulations were performed by more than 20 independent groups to assess current CFD techniques. The primary goal of this article is to summarize the FDA initiative and to report recent findings from the benchmark blood pump model study. Discrepancies between CFD predicted velocities and those measured using particle image velocimetry most often occurred in regions of flow separation (e.g., downstream of the nozzle throat, and in the pump exit diffuser). For the six pump test conditions, 57% of the CFD predictions of pressure head were within one standard deviation of the mean measured values. Notably, only 37% of all CFD submissions contained hemolysis predictions. This project aided in the development of an FDA Guidance Document on factors to consider when reporting computational studies in medical device regulatory submissions. There is an accompanying podcast available for this article. Please visit the journal's Web site (www.asaiojournal.com) to listen.

MeSH 主题词
Benchmarking Heart-Assist Devices Humans Hydrodynamics Models, Theoretical Rheology United States United States Food and Drug Administration
作者与单位
共 11 位作者,点击展开单位 / ORCID
Malinauskas Richard A
From the *Center for Devices and Radiological Health, Office of Science and Engineering Laboratories, Food and Drug Administration, Silver Spring, Maryland; †Department of Biomedical Engineering, Rochester Institute of Technology, Rochester, New York; ‡Department of Cardiovascular Engineering, RWTH Aachen University, Aachen, Germany; and §Department of Biomedical Engineering, Pennsylvania State University, University Park, Pennsylvania.
Hariharan Prasanna
Day Steven W
Herbertson Luke H
Buesen Martin
Steinseifer Ulrich
Aycock Kenneth I
Good Bryan C
Deutsch Steven
Manning Keefe B
Craven Brent A
Article Info
Journal
ASAIO journal (American Society for Artificial Internal Organs : 1992)
Abbr.
ASAIO J
ISSN
1538-943X
Published
2017-00-00
页码
150-160
Language
English
Country/Region
United States
NLM ID
9204109
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]