Biomechanical Evaluation of a Novel Loop Retention Mechanism for Cortical Graft Fixation in ACL Reconstruction.

Orthopaedic journal of sports medicine ·Vol. 8 ·No. 2 ·2020-02-00 ·页码 2325967120904322

Götschi T, Rosenberg G, Li X, Zhang C, Bachmann E, Snedeker JG, Fucentese SF

Abstract

Implant fixation by means of a cortical fixation device (CFD) has become a routine procedure in anterior cruciate ligament reconstruction. There is no clear consensus whether adjustable-length CFDs are more susceptible to loop lengthening when compared with pretied fixed-length CFDs. To assess biomechanical performance measures of 3 types of CFDs when subjected to various loading protocols. Controlled laboratory study. Three types of CFDs underwent biomechanical testing: 1 fixed length and 2 adjustable length. One of the adjustable-length devices is based on the so-called finger trap mechanism, and the other is based on a modified sling lock mechanism. A device-only test of 5000 cycles (n = 8 per group) and a tendon-device test of 1000 cycles (n = 8 per group) with lower and upper force limits of 50 and 250 N, respectively, were applied, followed by ramp-to-failure testing. Adjustable-length devices then underwent further cyclic testing with complete loop unloading (n = 5 per group) at each cycle, as well as fatigue testing (n = 3 per group) over a total of 1 million cycles. Derived mechanical parameters were compared among the devices for statistical significance using Kruskal-Wallis analysis of variance followed by post hoc Mann-Whitney U testing with Bonferroni correction. All CFDs showed elongation <2 mm after 5000 cycles when tested in an isolated manner and withstood ultimate tensile forces in excess of estimated peak in vivo forces. In both device-only and tendon-device tests, differences in cyclic performance were found among the devices, favoring adjustable-length fixation devices over the fixed-length device. Completely unloading the suspension loops, however, led to excessive loop lengthening of the finger trap device, whereas the modified sling lock device remained stable throughout the test. The fixed-length device displayed superior ultimate strength over both adjustable-length devices. Both adjustable-length devices showed adequate fatigue behavior during high-cyclic testing. All tested devices successfully prevented critical construct elongation when tested with constant tension and withstood ultimate loads in excess of estimated in vivo forces during the rehabilitation phase. The finger trap device gradually lengthened excessively when completely unloaded during cyclic testing. Critical loop lengthening may occur if adjustable-length devices based on the finger trap mechanism are repeatedly unloaded in situ.

Keywords
ACL reconstruction adjustable button biomechanical testing cortical fixation cyclic loading
作者与单位
共 7 位作者,点击展开单位 / ORCID
Götschi Tobias
Institute for Biomechanics, ETH Zurich, Zurich, Switzerland. | Department of Orthopedics, Balgrist University Hospital, Zurich, Switzerland.
Rosenberg George
Institute for Biomechanics, ETH Zurich, Zurich, Switzerland.
Li Xiang
Institute for Biomechanics, ETH Zurich, Zurich, Switzerland.
Zhang Chen
Institute for Biomechanics, ETH Zurich, Zurich, Switzerland. | Department of Orthopedics, Second Affiliated Hospital of Xi'an Jiaotong University, Xi'an, China.
Bachmann Elias
Institute for Biomechanics, ETH Zurich, Zurich, Switzerland. | Department of Orthopedics, Balgrist University Hospital, Zurich, Switzerland.
Snedeker Jess G
Institute for Biomechanics, ETH Zurich, Zurich, Switzerland. | Department of Orthopedics, Balgrist University Hospital, Zurich, Switzerland.
Fucentese Sandro F
Department of Orthopedics, Second Affiliated Hospital of Xi'an Jiaotong University, Xi'an, China.
Article Info
Journal
Orthopaedic journal of sports medicine
Abbr.
Orthop J Sports Med
ISSN
2325-9671
Published
2020-02-00
电子出版
2020-00-25
页码
2325967120904322
Language
English
Country/Region
United States
NLM ID
101620522
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]