Home LiteratureArticle Details
PMID: 34243316 Published · ppublish English Journal Article

Fluid dynamics analysis and experimental study for solar radiation error correction of sounding humidity sensor.

The Review of scientific instruments ·Vol. 92 ·No. 5 ·2021-05-01 ·页码 055010

Zhang J, Xie X, Liu Q, Li M, Mao X

Abstract

In order to correct the solar radiation error of relative humidity, the mainstream capacitive sounding humidity sensor HC103M2 is selected and investigated by simulation analysis and experimental verification. First, the basic theories for solar radiation error and sensor error itself are elaborated, and simulational and experimental platforms are introduced. The computational fluid dynamics (CFD) method is utilized to theoretically investigate the dry error of the humidity sensor caused by solar radiation heating, which is related to radiation intensity, altitude, and solar elevation angle as well as reflectivity, thickness, and shape of the shield. Then, in order to verify the accuracy of the simulation, an experimental platform including a humidity sensor and two temperature sensors to measure the solar radiation heating is built to analyze the relative error of humidity obtained by the CFD simulation and experiment. It is found that their maximum deviation is 3.30% and the average error is 1.94%, which indicates that the calculation using the CFD method is accurate and feasible. In order to easily and operationally predict the solar radiation heating of the humidity sensor, a back propagation (BP) neural network fusion algorithm based on three inputs of radiation intensity, air pressure, and solar elevation angle is proposed. Compared with the solar radiation heating obtained by CFD simulation, the maximum absolute error is about 0.2 K, and the relative error of humidity is about ±1.30%. Finally, a case of vertical humidity profile correction considering the temperature-sensitive error of HC103M2 is analyzed. The response time of sensor measurement and the airflow into the shield are discussed as well. The corrected results after taking solar radiation heating into account are more similar to those measured by RS92 and cryogenic frost point hygrometer (CFH). This result shows that the prediction model is accurate, which may be applied to correct the dry error and further improve the accuracy of humidity measurement.

作者与单位
共 5 位作者,点击展开单位 / ORCID
Zhang Jiahong ORCID
Jiangsu Key Laboratory of Meteorological Observation and Information Processing, Nanjing University of Information Science and Technology, Nanjing 210044, China.
Xie Xiaolu ORCID
Jiangsu Key Laboratory of Meteorological Observation and Information Processing, Nanjing University of Information Science and Technology, Nanjing 210044, China.
Liu Qingquan ORCID
Jiangsu Key Laboratory of Meteorological Observation and Information Processing, Nanjing University of Information Science and Technology, Nanjing 210044, China.
Li Min
Jiangsu Key Laboratory of Meteorological Observation and Information Processing, Nanjing University of Information Science and Technology, Nanjing 210044, China.
Mao Xiaoli
Jiangsu Key Laboratory of Meteorological Observation and Information Processing, Nanjing University of Information Science and Technology, Nanjing 210044, China.
Article Info
Journal
The Review of scientific instruments
Abbr.
Rev Sci Instrum
ISSN
1089-7623
Published
2021-05-01
页码
055010
Language
English
Country/Region
United States
NLM ID
0405571
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]