Home LiteratureArticle Details
PMID: 3662001 Published · ppublish English Journal Article Research Support, U.S. Gov't, P.H.S.

Microsequence analysis of peptides and proteins. VI. A continuous flow reactor for sample concentration and sequence analysis.

Analytical biochemistry ·Vol. 163 ·No. 2 ·1987-06-00 ·Pages 517-29

Shively JE, Miller P, Ronk M

Abstract

We have designed and tested a continuous flow reactor (CFR) for microsequence analysis of peptides and proteins. The CFR forms the site for immobilization of the peptide or protein substrate and automated Edman chemistry. The CFR was constructed from 0.125-in.-o.d., 0.0625-in.-i.d. Teflon tubing (length 2-3 cm) containing 5-10 mg of Polybrene-coated, spherical, porous silica (100-200-micron particle size). The silica is retained in the CFR with porous Teflon filters (Zitex) at the bed bottom and optionally at the bed top. The i.d. of the CFR was selected for a tight press fit when 0.0625-in.-o.d. Teflon lines are inserted at the top and bottom of the CFR. This design allows the replacement of the existing cartridge/glass fiber disk found in conventional microsequencers with a CFR with a minimal amount of changes. The advantages of the CFR over the previous design include a lower background or noise level and no need to precycle Polybrene before sample application, and the entire unit is inexpensive and therefore disposable. We believe that the decrease in noise, especially the decrease in the commonly observed diphenylthiourea peak, is due to the more direct flow path and relative absence of unswept area in the CFR. Several standard peptides and proteins were sequenced in the CFR to demonstrate the improved results. A direct comparison to the cartridge/glass fiber disk design demonstrated less background and higher initial and repetitive yields for the CFR. An additional advantage is the ability to directly concentrate samples on CFRs containing reverse-phase packing. We have successfully concentrated 1.0-ml samples (200 pmol) onto 5 mg of octyldecylsilyl-derivatized silica in yields of 95-100%. The resulting samples were microsequenced after addition of Polybrene-coated silica to the CFR with high initial and repetitive yields. This methodology promises to improve sample handling and microsequence analysis of low picomole amounts of peptides and proteins.

MeSH Terms
Amino Acid Sequence Chromatography, High Pressure Liquid/instrumentation Peptides Proteins
Chemicals
Peptides Proteins
Authors & Affiliations
3 authors, click to expand affiliations / ORCID
Shively J E
Division of Immunology, Beckman Research Institute of the City of Hope, Duarte, California 91010.
Miller P
Ronk M
Article Info
Journal
Analytical biochemistry
Abbr.
Anal Biochem
ISSN
0003-2697
Published
1987-06-00
Pages
517-29
Language
English
Region
United States
NLM ID
0370535
Subset
IM
Grants
NIADDK NIH HHS · AM 33155 · United States
NCI NIH HHS · CA 37808 · United States
NICHD NIH HHS · HD 14900 · United States
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]