PMID: 36745955 Published · ppublish English

A mathematical model for assessing shear induced bleeding risk.

Li Y, Wang H, Xi Y, Sun A, Wang L, Deng X, Chen Z, Fan Y

Abstract

The objective of this study is to develop a bleeding risk model for assessing device-induced bleeding risk in patients supported with blood contact medical devices (BCMDs). The mathematical model for evaluating bleeding risk considers the effects of shear stress on von Willebrand factor (vWF) unfolding, high molecular weight multimers-vWF (HMWM-vWF) degradation, platelet activation and receptor shedding and platelet-vWF binding ability. Functions of the effect of shear stress on the above factors are fitted/employed and solved by the Eulerian transport equation. An axial flow-through Couette device and two clinical VADs which are HeartWare Ventricular Assist Device (HVAD) and HeartMate II (HM II) blood pump were employed to perform the simulation to evaluate platelet receptor shedding (GPIbα and GPIIb/IIIa), loss of HWMW-vWF, platelet-vWF binding ability and bleeding risk for validating the accuracy of our model. The platelet-vWF binding ability after being subjected to high shear region in the axial flow-through Couette device predicted by our bleeding model was highly consistent with reported experimental data. As indicated by our CFD simulation results in the axial flow-through Couette device, it can find that an increase in shear stress led to a decrease in the adhesion ability of platelets on vWF, while the binding ability of vWF with platelets first increase and then decrease as shear stress elevates gradually beyond a threshold. The factor of exposure time can enhance the effect of shear stress. Additionally, the shear-induced bleeding risk predicted by our model increases with increasing shear stress and exposure time in an axial flow-through Couette device. As indicated by our numerical model, the bleeding risk in HVAD was higher than HMII, which is highly consistent with the meta-analysis based on clinical statistics. Our simulation investigations in these two clinical VADs also found that HVAD caused a higher rate of platelet receptor shedding and lower damage to HWMW-vWF than HeartMate II. The high shear stress generated in the narrow and turbulent regions of both VADs was the underlying cause of device-induced bleeding. In this study, the shear-induced bleeding risk predicted by our bleeding model in axial flow-through Couette device and two clinical VADs is consistent or highly correlated with experimental and clinical findings, which proves the accuracy of our bleeding model. Our bleeding model can be used to aid the development of new BCMDs with improved functional characteristics and biocompatibility, and help to reduce risk of device-induced adverse events in patients.

Keywords
Bleeding Blood contact medical devices Platelet receptors Shear stress VAD vWF
MeSH 主题词
Humans von Willebrand Factor/analysis Hemorrhage/etiology,metabolism Platelet Activation Blood Platelets/chemistry,metabolism Heart-Assist Devices Stress, Mechanical Models, Theoretical
Article Info
Journal
Computer methods and programs in biomedicine
Abbr.
Comput Methods Programs Biomed
ISSN
1872-7565
Published
2023-04-00
Language
English
Country/Region
Ireland
NLM ID
8506513
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]