Optimization of 4D Flow MRI Spatial and Temporal Resolution for Examining Complex Hemodynamics in the Carotid Artery Bifurcation.

Cardiovascular engineering and technology ·Vol. 14 ·No. 3 ·2023-00-00 ·页码 476-488

El Sayed R, Sharifi A, Park CC, Haussen DC, Allen JW, Oshinski JN

Abstract

Three-dimensional, ECG-gated, time-resolved, three-directional, velocity-encoded phase-contrast MRI (4D flow MRI) has been applied extensively to measure blood velocity in great vessels but has been much less used in diseased carotid arteries. Carotid artery webs (CaW) are non-inflammatory intraluminal shelf-like projections into the internal carotid artery (ICA) bulb that are associated with complex flow and cryptogenic stroke. Optimize 4D flow MRI for measuring the velocity field of complex flow in the carotid artery bifurcation model that contains a CaW. A 3D printed phantom model created from computed tomography angiography (CTA) of a subject with CaW was placed in a pulsatile flow loop within the MRI scanner. 4D Flow MRI images of the phantom were acquired with five different spatial resolutions (0.50-2.00  mm3) and four different temporal resolutions (23-96 ms) and compared to a computational fluid dynamics (CFD) solution of the flow field as a reference. We examined four planes perpendicular to the vessel centerline, one in the common carotid artery (CCA) and three in the internal carotid artery (ICA) where complex flow was expected. At these four planes pixel-by-pixel velocity values, flow, and time average wall shear stress (TAWSS) were compared between 4D flow MRI and CFD. An optimized 4D flow MRI protocol will provide a good correlation with CFD velocity and TAWSS values in areas of complex flow within a clinically feasible scan time (~ 10 min). Spatial resolution affected the velocity values, time average flow, and TAWSS measurements. Qualitatively, a spatial resolution of 0.50  mm3 resulted in higher noise, while a lower spatial resolution of 1.50-2.00  mm3 did not adequately resolve the velocity profile. Isotropic spatial resolutions of 0.50-1.00  mm3 showed no significant difference in total flow compared to CFD. Pixel-by-pixel velocity correlation coefficients between 4D flow MRI and CFD were > 0.75 for 0.50-1.00  mm3 but were < 0.5 for 1.50 and 2.00  mm3. Regional TAWSS values determined from 4D flow MRI were generally lower than CFD and decreased at lower spatial resolutions (larger pixel sizes). TAWSS differences between 4D flow and CFD were not statistically significant at spatial resolutions of 0.50-1.00  mm3 but were different at 1.50 and 2.00 mm3. Differences in temporal resolution only affected the flow values when temporal resolution was > 48.4 ms; temporal resolution did not affect TAWSS values. A spatial resolution of 0.74-1.00  mm3 and a temporal resolution of 23-48 ms (1-2 k-space segments) provides a 4D flow MRI protocol capable of imaging velocity and TAWSS in regions of complex flow within the carotid bifurcation at a clinically acceptable scan time.

Keywords
4D flow MRI Carotid web Phantom model Spatial resolution Temporal resolution
MeSH 主题词
Magnetic Resonance Imaging/methods Hemodynamics Carotid Arteries/diagnostic imaging Pulsatile Flow Stress, Mechanical Blood Flow Velocity
作者与单位
共 6 位作者,点击展开单位 / ORCID
El Sayed Retta ORCID
Department of Biomedical Engineering, The Wallace H. Coulter, Emory University and Georgia Institute of Technology, Atlanta, GA, USA.
Sharifi Alireza
Department of Radiology & Imaging Sciences, Emory University, 1364 Clifton Rd, Atlanta, GA, 30322, USA.
Park Charlie C
Department of Radiology & Imaging Sciences, Emory University, 1364 Clifton Rd, Atlanta, GA, 30322, USA.
Haussen Diogo C
Department of Neurology, Emory University, Atlanta, GA, USA.
Allen Jason W
Department of Biomedical Engineering, The Wallace H. Coulter, Emory University and Georgia Institute of Technology, Atlanta, GA, USA. | Department of Radiology & Imaging Sciences, Emory University, 1364 Clifton Rd, Atlanta, GA, 30322, USA. | Department of Neurology, Emory University, Atlanta, GA, USA.
Oshinski John N ORCID
Department of Biomedical Engineering, The Wallace H. Coulter, Emory University and Georgia Institute of Technology, Atlanta, GA, USA. [email protected]. | Department of Radiology & Imaging Sciences, Emory University, 1364 Clifton Rd, Atlanta, GA, 30322, USA. [email protected]. | Department of Neurology, Emory University, Atlanta, GA, USA. [email protected].
Article Info
Journal
Cardiovascular engineering and technology
Abbr.
Cardiovasc Eng Technol
ISSN
1869-4098
Corresponding email
Published
2023-00-00
电子出版
2023-00-08
页码
476-488
Language
English
Country/Region
United States
NLM ID
101531846
基金资助
NIBIB NIH HHS · R01 EB027774 · United States
NINDS NIH HHS · R21 NS114603 · United States
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]