Impact of Partial Echo on 4D Flow MRI: The Insight From Synthetic MRI.

Magnetic resonance in medicine ·Vol. 95 ·No. 6 ·2026-06-00

Garreau M, Puiseux T, Moreno R, Toupin S, Giese D, Nicoud F, Mendez S

Abstract

The aim of this study is to investigate the impact of the partial echo on 4D flow MRI sequences thanks to in silico coupled MRI-CFD (computational fluid dynamics) simulations. Two sequences are studied: one with a full echo (FE) and another using partial echo (PE) with an echo symmetry fraction of 0.75. MRI-CFD simulations are conducted on an in silico pulsatile flow phantom for a sinusoidal inflow and a physiological inflow typical of the ascending aorta. For both inflow signals, PE-based simulations exhibited better compliance with their matching CFD simulations compared to the FE ones. The reduction of flow misregistration artifacts achieved through the use of PE appears to be more beneficial than the drawback of incomplete k-space filling. The MRI-CFD framework presented in this study appears as a useful tool to investigate the design of MRI sequences and to stratify its different sources of errors.

Keywords
4D flow computational fluid dynamics (CFD) numerical simulation partial echo (PE)
Article Info
Journal
Magnetic resonance in medicine
Abbr.
Magn Reson Med
ISSN
1522-2594
Published
2026-06-00
Language
English
Country/Region
United States
NLM ID
8505245
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]