PMID: 4874462 Published · ppublish English Journal Article

Microbiological studies on the performance of a laminar airflow biological cabinet.

Applied microbiology ·Vol. 16 ·No. 7 ·1968-07-00 ·Pages 1086-92

McDade JJ, Sabel FL, Akers RL, Walker RJ

Abstract

Engineering and microbiological tests indicated that a typical, commercial laminar airflow cabinet was not effective in providing either product protection or agent containment. The cabinet was modified and tested through a series of alternate configurations to establish a set of design criteria. A mock-up cabinet was developed from these design criteria. The mock-up unit was evaluated for efficiency in providing both product protection and agent containment. In these evaluations, challenge methods were developed to simulate normal, in-use laboratory operations. Controlled bacterial or viral aerosol challenges were used at higher than normal levels to provide stringent test conditions. Test results indicated that the mock-up unit was considerably better in preventing agent penetration (0.1 to 0.2 particles per 100 ft(3) of air) than the commercial cabinet (5 to 6 particles per 100 ft(3) of air) during product protection tests. Similarly, agent containment was considerably better in the new cabinet (particle escape of 2 to 3 per 100 ft(3) of air at only one of the five test sites) than in the commercial cabinet (particle escape of 2 to 14 per 100 ft(3) of air at three of the five test sites).

MeSH Terms
Aerosols Bacteriophages Equipment and Supplies Laboratories Laboratory Infection/prevention & control Microbiology Serratia marcescens/isolation & purification
Chemicals
Aerosols
Authors & Affiliations
4 authors, click to expand affiliations / ORCID
McDade J J
Sabel F L
Akers R L
Walker R J
References (5)
5 references, click to expand
  1. Ultraclean operating room.
    Am J Surg. 1966 Nov;112(5):681-5 PMID: 5923397
  2. Comparative levels and types of microbial contamination detected in industrial clean rooms.
    Appl Microbiol. 1966 Jul;14(4):539-51 PMID: 5954479
  3. Use of laminar air-flow equipment in microbiology.
    Appl Microbiol. 1968 Jan;16(1):182-3 PMID: 5636465
  4. Evaluation of laboratory fume hoods.
    Am Ind Hyg Assoc Q. 1954 Sep;15(3):195-202 PMID: 13188836
  5. A protective cabinet for handling infective material in the laboratory.
    J Clin Pathol. 1957 Nov;10(4):400-2 PMID: 13481133
Article Info
Journal
Applied microbiology
Abbr.
Appl Microbiol
ISSN
0003-6919
Published
1968-07-00
Pages
1086-92
Language
English
Region
United States
NLM ID
7605802
PMCID
PMC547593
Subset
IM
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]