Abstract
The N/F system (Flow Laboratories, Inc. Rockville, Md) and the Oxi/Ferm Tube, (Roche Diagnostics, Div. Hoffman-La Roche, Inc. Nutley, N.J.) were evaluated in parallel for identification of infrequently encountered nonfermentative gram-negative bacilli and oxidase-positive, fermentative gram-negative bacilli selected from fresh clinical isolates and stock cultures. When compared with conventional media, the Flow N/F system correctly identified 97.7% (86 or 88) of all strains tested. No organisms were misidentified, but this system failed to identify two strains. Six supplemental tests were needed for the complete identification of 16 strains (18%); however, correct results were obtained within 48 h for 85% of the isolates. The Oxi/Ferm method correctly identified 87.5% of the isolates; 7% were incorrectly identified, and the method failed to identify five strains. Seventeen supplemental tests were required to identify 64 strains (73%). Complete identification within 48 h was obtained for 60% of the organisms tested.
MeSH Terms
Bacteria/classification,enzymology
Bacteriological Techniques
Fermentation
Oxidoreductases/analysis
Chemicals
Oxidoreductases
Authors & Affiliations
3 authors, click to expand affiliations / ORCID
Koestenblatt E K
Larone D H
Pavletich K J
References (9)
9 references, click to expand
-
Evaluation of two test-kits--API and Oxi Ferm tube--for identification of oxidative-fermentative Gram-negative rods.
Med Microbiol Immunol. 1977 Jul 18;163(2):93-7
PMID: 331064
-
Evaluation of the oxi/ferm tube system with selected Gram-negative bacteria.
J Clin Microbiol. 1977 Dec;6(6):559-66
PMID: 338624
-
Evaluation of two rapid methods for identification of commonly encountered nonfermenting or oxidase-positive, Gram-negative rods.
J Clin Microbiol. 1977 Dec;6(6):605-9
PMID: 338626
-
Evaluation of the Oxi/Ferm tube system for identification of nonfermentative Gram-negative bacilli.
J Clin Microbiol. 1978 Jun;7(6):533-8
PMID: 353066
-
Accuracy and reproducibility of the Oxi/Ferm system in identifying a select group of unusual gram-negative bacilli.
J Clin Microbiol. 1979 Feb;9(2):180-5
PMID: 372222
-
Comparison of the API 20E and Oxi/Ferm systems in identification of nonfermentative and oxidase-positive fermentative bacteria.
J Clin Microbiol. 1979 Feb;9(2):220-6
PMID: 372226
-
Rapid identification of nonfermentative gram-negative rods by the Corning N/F system.
J Clin Microbiol. 1979 Feb;9(2):239-43
PMID: 107190
-
Comparison of the API 20E and Corning N/F systems for identification of nonfermentative gram-negative rods.
J Clin Microbiol. 1979 Aug;10(2):175-9
PMID: 389947
-
Four methods for identification of gram-negative nonfermenting rods: organisms more commonly encountered in clinical specimens.
J Clin Microbiol. 1980 Aug;12(2):271-8
PMID: 7014610