Home LiteratureArticle Details
PMID: 736424 Published · ppublish English Journal Article

Auditory prostheses research with multiple channel intracochlear stimulation in man.

The Annals of otology, rhinology, and laryngology ·Vol. 87 ·No. 6 Pt 2 ·1978-00-00 ·Pages 1-39

Eddington DK, Dobelle WH, Brackmann DE, Mladejovsky MG, Parkin JL

Abstract

Although single-channel electrode arrays implanted in the scala tympani of deaf patients are useful as an aid to lip reading and for distinguishing some environmental sounds, they do not transmit intelligible speech. However, multichannel electrode arrays, which take advantage of the cochlea's tonotopic organization, may be capable of generating the complex patterns of neural activity necessary for speech discrimination. In this study, multichannel electrodes were implanted in the cochleas of four volunteers, with access to the connecting wires made through the skin via a percutaneous connector. The major portion of the data presented is from two of these subjects: one has been bilaterally deaf since birth and the other has been unilaterally deaf for 15 years. Preliminary results of experiments with two more recently implanted subjects are described as well as experiments with a fifth volunteer who was implanted with five electrodes by House in 1969. Data on pitch and loudness discrimination as well as the effects of stimulation parameters on threshold, impedance, and electrode interaction are presented. Place pitch and periodicity pitch were observed in all five volunteers. The results of pitch-matching experiments with the unilaterally deaf volunteer were consistent with tonotopic maps of the cochlea, and experiments indicated that a pitch continuum may be achieved by combining place and periodicity pitch modulation. Preliminary experiments in tune recognition with one subject demonstrate his ability to recognize simple melodies based on periodicity pitch cues. These results, coupled with the finding that subjective sensations remain stable over the long-term, support the feasibility of providing artificial hearing with a multichannel cochlear stimulation system.

MeSH Terms
Acoustic Stimulation/instrumentation Auditory Threshold Biocompatible Materials Cochlea/physiology,surgery Computers Deafness/rehabilitation Electric Conductivity Electric Stimulation/instrumentation Electrodes, Implanted Electronics, Medical/instrumentation Forecasting Hearing Aids Hearing Loss/rehabilitation Humans Informed Consent Loudness Perception Male Middle Aged Pitch Discrimination Prostheses and Implants Prosthesis Design Scala Tympani/physiology,surgery Sensation Speech Perception Time Factors
Chemicals
Biocompatible Materials
Authors & Affiliations
5 authors, click to expand affiliations / ORCID
Eddington D K
Dobelle W H
Brackmann D E
Mladejovsky M G
Parkin J L
Article Info
Journal
The Annals of otology, rhinology, and laryngology
Abbr.
Ann Otol Rhinol Laryngol
ISSN
0003-4894
Published
1978-00-00
Pages
1-39
Language
English
Region
United States
NLM ID
0407300
Subset
IM
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]