Home LiteratureArticle Details
PMID: 7419705 Published · ppublish English Journal Article

Clinical evaluation of automated antibiotic susceptibility testing with the MS-2 system.

Journal of clinical microbiology ·Vol. 12 ·No. 4 ·1980-10-00 ·Pages 527-32

Barnes WG, Green LR, Talley RL

Abstract

The MS-2 (Abbott Laboratories) system for automated antimicrobial susceptibility testing was evaluated for both accuracy and general utility in our clinical laboratory. A total of 984 fresh clinical bacterial isolates (745 gram-negative, 239 gram-positive) were tested with the MS-2 system, and results were compared directly with those from a conventional agar disk diffusion method. Discrepancies between the two methods were categorized as very major, major, and minor. For gram-positive isolates, full accord (all discrepancies considered) was 91.6%, and essential accord (minor discrepancies not included) was 96.2%. With gram-negative isolates, full accord was found to be 93.9%, with essential accord of 97.9%. Aggrement as a function both of organism group and of antimicrobial agent was determined. Full accord of 90% or more was found for all major organism groups tested, with the exception of enterococci, where discrepant results between the two methods were observed. Mean test time for all isolates tested was 4.3 h. The MS-2 was found to be an accurate and highly automated instrument which required minimal technician time and was readily adaptable to work flow in our clinical laboratory.

MeSH Terms
Anti-Bacterial Agents/pharmacology Autoanalysis Enterobacteriaceae/drug effects Evaluation Studies as Topic Microbial Sensitivity Tests/methods Pseudomonas/drug effects Staphylococcus/drug effects Streptococcus/drug effects
Chemicals
Anti-Bacterial Agents
Authors & Affiliations
3 authors, click to expand affiliations / ORCID
Barnes W G
Green L R
Talley R L
References (3)
3 references, click to expand
  1. Antibiotic susceptibility testing by a standardized single disk method.
    Am J Clin Pathol. 1966 Apr;45(4):493-6 PMID: 5325707
  2. Laboratory evaluation of a rapid, automatic susceptibility testing system: report of a collaborative study.
    Antimicrob Agents Chemother. 1975 Apr;7(4):466-80 PMID: 1147582
  3. Clinical laboratory evaluation of the Abbott MS-2 automated antimicrobial susceptibility testing system: report of a collaborative study.
    J Clin Microbiol. 1980 Sep;12(3):375-90 PMID: 6783678
Article Info
Journal
Journal of clinical microbiology
Abbr.
J Clin Microbiol
ISSN
0095-1137
Published
1980-10-00
Pages
527-32
Language
English
Region
United States
NLM ID
7505564
PMCID
PMC273629
Subset
IM
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]