Home LiteratureArticle Details
PMID: 7513408 Published · ppublish English Journal Article Research Support, U.S. Gov't, Non-P.H.S. Research Support, U.S. Gov't, P.H.S. Review

Why do some patients have high defibrillation thresholds at defibrillator implantation? Answers from basic research.

Pacing and clinical electrophysiology : PACE ·Vol. 17 ·No. 2 ·1994-02-00 ·Pages 222-39

Hillsley RE, Wharton JM, Cates AW, Wolf PD, Ideker RE

Abstract

Implantable cardioverter defibrillators reduce the risk of sudden cardiac death in patients with ventricular tachyarrhythmias. However, for the few patients with unacceptably high defibrillation thresholds at implantation the risk of sudden death may remain high. If a small number of defibrillation attempts are used to determine a defibrillation threshold, then a high defibrillation threshold may occur in some patients due to the probabilistic nature of defibrillation: a small percentage of shocks will fail even at optimal shock strengths. Basic investigations have suggested mechanisms for high defibrillation thresholds in other patients. The extracellular potential gradients produced by a shock correlate with ability to defibrillate and may be used to classify mechanisms for high defibrillation thresholds. Computerized mapping studies have demonstrated that extracellular potential gradient fields produced by defibrillation shocks are uneven with high gradient areas close to the electrodes and low gradient areas distant from the electrodes. A high defibrillation threshold may occur because: (1) a shock creates a subthreshold potential gradient in the low gradient areas; (2) a patient has a higher minimum potential gradient threshold than other patients; or (3) a shock leads to refibrillation in the high gradient areas. This article reviews experimental evidence to support each of these three possibilities then suggests experimental and clinical investigations that may clarify the causes of high defibrillation thresholds in patients.

MeSH Terms
Defibrillators, Implantable Electric Conductivity Electric Countershock/methods Electric Impedance Heart/physiopathology Humans Tachycardia, Ventricular/physiopathology,therapy
Authors & Affiliations
5 authors, click to expand affiliations / ORCID
Hillsley R E
Department of Medicine, Duke University, Durham, North Carolina 27710.
Wharton J M
Cates A W
Wolf P D
Ideker R E
Article Info
Journal
Pacing and clinical electrophysiology : PACE
Abbr.
Pacing Clin Electrophysiol
ISSN
0147-8389
Published
1994-02-00
Pages
222-39
Language
English
Region
United States
NLM ID
7803944
Subset
IM
Grants
NHLBI NIH HHS · HL-28429 · United States
NHLBI NIH HHS · HL-42760 · United States
NHLBI NIH HHS · HL-44066 · United States
Corrections
CommentIn
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]