The effect of cardiac compression on defibrillation efficacy and the upper limit of vulnerability.

Journal of cardiovascular electrophysiology ·Vol. 6 ·No. 5 ·1995-05-00 ·Pages 368-78

Idriss SF, Anstadt MP, Anstadt GL, Ideker RE

Abstract

We determined the effects of decreasing the ventricular blood volume and altering cardiac geometry on defibrillation, the upper limit of vulnerability (ULV), and the relationship between them. In six pigs, fibrillation/defibrillation trials were performed with a left ventricular apex patch to a superior vena cava catheter electrode configuration and a biphasic waveform. Thirty trials each were performed on a compressed versus noncompressed (normal) heart. Compression was achieved using direct mechanical ventricular actuation. Dose-response curves were constructed, and the 50% probability points (ED50) were compared for leading edge voltage (LEV), leading edge current (LEI), and total energy (TE). In another 12 pigs, triplicate defibrillation thresholds (DFTs) and ULVs were determined for each heart state. The T wave was scanned with shocks in 10-msec steps for determining the ULV. Compression resulted in decreased ED50s for LEV (delta = 138 +/- 77 V, P < 0.05, mean +/- SD), LEI (delta = 1.57 +/- 0.7 A, P < 0.05), and TE (delta = 4.9 +/- 3.6 J, P < 0.05) compared to normal. In the second study, compression significantly reduced DFT (P < 0.02) and ULV (P < 0.02) for LEV, LEI, and TE compared to normal. The ULV tended to be lower than the DFT for the normal heart state (delta = 23 +/- 46 V LEV: P = NS). However, the ULV was significantly greater than the DFT for the compressed heart state (delta = 19 +/- 25 V LEV; P < 0.03). Shock delivery during cardiac compression improves defibrillation efficacy. Additionally, cardiac compression decreases both DFT and ULV, which supports the ULV hypothesis of defibrillation. Finally, maintaining the heart's geometric and volumetric state during ULV testing in paced rhythm and DFT testing in ventricular fibrillation moves the ULV higher than the DFT-the position predicted by the ULV hypothesis for defibrillation.

MeSH Terms
Animals Cardiac Volume/physiology Electric Countershock Electrodes Electroshock Heart/anatomy & histology,physiology In Vitro Techniques Swine Ventricular Function
Authors & Affiliations
4 authors, click to expand affiliations / ORCID
Idriss S F
Department of Pathology, Duke University Medical Center, Durham, North Carolina 27710, USA.
Anstadt M P
Anstadt G L
Ideker R E
Article Info
Journal
Journal of cardiovascular electrophysiology
Abbr.
J Cardiovasc Electrophysiol
ISSN
1045-3873
Published
1995-05-00
Pages
368-78
Language
English
Region
United States
NLM ID
9010756
Subset
IM
Grants
NHLBI NIH HHS · HL-42760 · United States
NHLBI NIH HHS · HL-44066 · United States
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]