Physical performance evaluation of the Toshiba GCA-9300A triple-headed system.

Kouris K, Clarke G A, Jarritt P H, Townsend C E, Thomas S N

摘要

The physical performance of the Toshiba GCA-9300A triple-headed SPECT system has been assessed. Using a water-filled cylinder containing 99mTc, the tomographic volume sensitivity was 33.8 and 34.8 kcps/(MBq/ml)/cm for the high-resolution, parallel-hole (HR-PH) collimator and the super high-resolution, lead fanbeam (SHR-FB) collimator, respectively, excluding the rotation time(s) during scanning when data are not acquired. The tomographic spatial resolution at the center, in air, with 132 mm radius of rotation was 10.2 and 7.8 mm FWHM with the HR-PH and SHR-FB collimators, respectively; in water it was 11.0 and 7.8 mm. Reconstructed relative activity concentrations were accurate for both collimator sets if attenuation correction was used. With the SHR-FB collimators, the average peak-to-valley ratio of five-line sources in water improved significantly when an asymmetric energy window was used. Using the three-dimensional Hoffman brain phantom which simulates a 4:1 grey matter-to-white matter ratio and the usual choice of acquisition and processing parameters for brain studies, the reconstructed grey matter-to-white matter ratio was only about 1.7 for total counts typically acquired in a 99mTc-HMPAO study (4.0 M counts) and only 2.3 for 40 M counts. There was a qualitative improvement with an asymmetric energy window.

文献信息
期刊
Journal of nuclear medicine : official publication, Society of Nuclear Medicine
期刊简称
J Nucl Med
发表日期
1993-11-09
收录日期
1993-11-09
更新日期
2016-11-23
语言
英语
国家/地区
United States
NLM ID
0217410
分析服务
分析服务

联系地址

山东省济南市章丘区文博路2号

齐鲁师范学院 genelibs生信实验室

山东省济南市高新区舜华路750号

大学科技园北区F座4单元2楼

电话: 0531-88819269

微信公众号

关注微信订阅号,实时查看信息,关注医学生物学动态。


商务邮箱

E-mail: [email protected]