Home LiteratureArticle Details
PMID: 7032421 Published · ppublish English Comparative Study Journal Article Research Support, U.S. Gov't, P.H.S.

Limulus amoebocyte lysate and direct sampling methods for surveillance of operating nebulizers.

Applied and environmental microbiology ·Vol. 42 ·No. 5 ·1981-11-00 ·Pages 850-5

Reinhardt DJ, Nabors W, Kennedy C, Malecka-Griggs B

Abstract

The Limulus amoebocyte lysate test for detection of endotoxin (Pyrogent; Mallinckrodt Chemical Co.) and the Easicult method (Orion Diagnostica) for detection of bacteria were compared with direct dilution sampling, a standardized technique for respiratory therapy surveillance previously developed in our laboratory. Tests of 206 reservoirs of nebulizers were done in three hospitals in Georgia. Forty-five percent of all reservoirs sampled were contaminated. Gram-negative, nonfermentative bacilli were the predominant contaminants. The results of the Limulus test and the Easicult system were in agreement with those of the direct dilution sampling tests approximately 84 and 90% of the time, respectively. Direct dilution of water samples onto blood agar plates was the most sensitive, reliable, and informative method for detecting viable bacteria. The Easicult and Limulus systems were sensitive enough to detect greater than or equal to 10(3) colony-forming units per ml. Positive Limulus tests and negative culture tests, reflecting detection of endotoxin but not of viable gram-negative bacteria, occurred in 20 of 206 (9.7%) instances. Positive cultures and negative Limulus tests were noted in 13 of 206 (6.8%) samplings. The Limulus test is a valuable procedure, for it can detect moderate-to-heavy microbial contamination within 1 h of testing and affords the opportunity to remove contaminated equipment from patients within minutes of a positive test result. These results demonstrate the potential value of the Easicult and Limulus tests for selective surveillance of operating nebulizers.

MeSH Terms
Acinetobacter/isolation & purification Alcaligenes/isolation & purification Bacteria/isolation & purification Bacteriological Techniques Endotoxins/analysis Limulus Test Pseudomonas/isolation & purification Respiratory Therapy/instrumentation Water Microbiology
Chemicals
Endotoxins
Authors & Affiliations
4 authors, click to expand affiliations / ORCID
Reinhardt D J
Nabors W
Kennedy C
Malecka-Griggs B
References (19)
19 references, click to expand
  1. Nebulization equipment. A potential source of infection in gram-negative pneumonias.
    Am J Dis Child. 1966 Apr;111(4):357-60 PMID: 5906045
  2. THE POTENTIAL ROLE OF INHALATION THERAPY EQUIPMENT IN NOSOCOMIAL PULMONARY INFECTION.
    J Clin Invest. 1965 May;44:831-9 PMID: 14276140
  3. Pseudomonas aeruginosa: growth in distilled water from hospitals.
    Science. 1971 Aug 27;173(3999):836-8 PMID: 4999114
  4. Factors affecting comparative resistance of naturally occurring and subcultured Pseudomonas aeruginosa to disinfectants.
    Appl Microbiol. 1972 May;23(5):863-9 PMID: 4624209
  5. Bacterial contamination of aerosols.
    Arch Intern Med. 1973 Jan;131(1):156-9 PMID: 4565025
  6. A practical method for monitoring bacterial contamination of inhalation therapy machines.
    Am Rev Respir Dis. 1972 Dec;106(6):920-2 PMID: 4641227
  7. Morphological, biochemical, and growth characteristics of pseudomonas cepacia from distilled water.
    Appl Microbiol. 1973 Mar;25(3):476-83 PMID: 4698864
  8. Preparation, sensitivity, and specificity of Limulus lysate for endotoxin assay.
    Appl Microbiol. 1973 Jul;26(1):43-8 PMID: 4199336
  9. Rapid detection of contaminated intravenous fluids using the Limulus in vitro endotoxin assay.
    Appl Microbiol. 1973 Oct;26(4):521-4 PMID: 4584590
  10. Microbiologic sampling of the hospital environment.
    Health Lab Sci. 1974 Apr;11(2):73-5 PMID: 4594405
  11. Sources of the slow-growing pigmented water bacteria.
    Health Lab Sci. 1976 Jan;13(1):5-10 PMID: 816754
  12. Rapid detection of bacterial endotoxins in drinking water and renovated wastewater.
    Appl Environ Microbiol. 1976 Sep;32(3):347-51 PMID: 791115
  13. Evaluation of a simple device for bacteriological sampling of respirator-generated aerosols.
    J Clin Microbiol. 1977 Feb;5(2):178-83 PMID: 403199
  14. Gram-negative water bacteria in hemodialysis systems.
    Health Lab Sci. 1975 Oct;12(4):321-34 PMID: 1236620
  15. The slow-growing pigmented water bacteria: problems and sources.
    Adv Appl Microbiol. 1978;23:155-71 PMID: 356536
  16. Comparison of Limulus assay, standard plate count, and total coliform count for microbiological assessment of renovated wastewater.
    Appl Environ Microbiol. 1979 May;37(5):928-31 PMID: 384901
  17. Selective nonroutine microbial surveillance of in-use hospital nebulizers by aerosol entrapment and direct sampling analyses of solutions in reservoirs.
    J Clin Microbiol. 1980 Aug;12(2):199-204 PMID: 6785304
  18. Survival and dissemination of bacteria in nebulizers and incubators.
    Am J Dis Child. 1967 Jul;114(1):13-20 PMID: 4952773
  19. Simple methods of bacteriologic sampling of nebulization equipment.
    Am Rev Respir Dis. 1966 Sep;94(3):450-3 PMID: 5331326
Article Info
Journal
Applied and environmental microbiology
Abbr.
Appl Environ Microbiol
ISSN
0099-2240
Published
1981-11-00
Pages
850-5
Language
English
Region
United States
NLM ID
7605801
PMCID
PMC244118
Subset
IM
Grants
NCRR NIH HHS · RR 09201 · United States
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: [email protected]